Peak table contains peak detection results and mass bias correction details
IsoCor perfroms mass bias correction according to the formula:\ $IR_{corrected}=IR_{measured}*K$
The correction factor K can be calculated with one of three equations that is selected by the user:\ Linear equation:\ $K_{SI/MI}=1+f(amu_{SI}-amu_{MI})$
Russel's equation:\ $K_{SI/MI}=(amu_{SI}/amu_{MI})^f$
Exponential equation:\ $K_{SI/MI}=e^{f*(amu_{SI}-amu_{MI})}$
After mass bias correction method is selected, the f values should be copy/pasted onto the placeholder. Typically, f values are calculated from continuously introduced standard, thus number of f values should be same as number of peaks (for example 3 repetitions files, each file has 3 peaks; thus total 9 peaks need to have 9 corresponding f values). Atomic weights of the isotopes are also added by the user for MI and SI on Import section.\ Then correction factor K is calculated by the App, and isotope ratio values are updated according to mass bias correction. Manually added f values and app-calculated K values are shown in Peak table.
Each peak can be selected as either standard or sample or be discarded. The selection of peak type is needed for delta calculation with standard-sample bracketing method.
For example, if file has three peaks, then, typically, the 1st and 3rd peaks are standard, and middle peak is sample.
To set the type of the peak:\ 1. click on one of the rows on Peak table\ 2. click change peak type button\ 3. select type of the peak from dropdown menu
After peak detection, each peak is indexed with a number. Peaks with same index are assigned with same type of peak. For example, when type of peak #1 on sample #1 is set to Standard, then peak #1 in all samples are assigned as Standard. Similarly, if type of peak #2 on sample #1 is set to Sample, then peak #1 in all samples are assigned as Sample. This is important to perform same calculation of delta in all samples.
Note!\ by default, when typical data with three peaks are uploaded, the first and third peaks are set to standard, and middle peak is set to sample. This default type can be changed following the steps described above. For delta calculation, the isotope ratio of standard is found by averaging isotope ratio values of first and third peaks.
Any scripts or data that you put into this service are public.
Add the following code to your website.
For more information on customizing the embed code, read Embedding Snippets.