lift_adat | R Documentation |
The SomaScan platform continually improves its technical processes between assay versions. The primary change of interest is content expansion, and other protocol changes may be implemented including: changing reagents, liquid handling equipment, and well volumes.
Table of SomaScan assay versions:
Version | Commercial Name | Size |
V4 | 5k | 5284 |
v4.1 | 7k | 7596 |
v5.0 | 11k | 11083 |
However, for a given analyte, these technical upgrades can result
in minute measurement signal differences,
requiring a calibration (aka "lifting" or "bridging") to bring RFUs into a
comparable signal space.
This is accomplished by applying an analyte-specific scalar,
a linear transformation, to each analyte RFU measurement (column).
If you have an annotations file (*.xlsx
) and wish to examine the
bridging scalars themselves, please see read_annotations()
.
Lifting between SomaScan versions no longer requires an
annotations file containing lifting scalars. We now enable users to pass
a bridge
parameter, indicating the direction of the bridge.
For example, to "lift" between 11k
-> 7k
, you must be acting on
SomaScan data in 11k
RFU space and would pass bridge = "11k_to_7k"
.
Likewise, 7k
-> 5k
requires bridge = "7k_to_5k"
.
Lastly, you may also lift directly from 11k
-> 5k
(aka "double-bridge") with bridge = "11k_to_5k"
.
See below for all options for the bridge
argument.
lift_adat(
adat,
bridge = c("11k_to_7k", "11k_to_5k", "7k_to_11k", "7k_to_5k", "5k_to_11k", "5k_to_7k"),
anno.tbl = deprecated()
)
is_lifted(adat)
adat |
A |
bridge |
The direction of the lift (i.e. bridge). |
anno.tbl |
Matched samples across assay versions are used to calculate bridging
scalars. For each analyte, this scalar is computed as the ratio of
population medians across assay versions.
Please see the lifting vignette
vignette("lifting-and-bridging", package = "SomaDataIO")
for more details.
lift_adat()
: A "lifted" soma_adat
object corresponding to
the scaling requested in the bridge
parameter. RFU values are
rounded to 1 decimal place to match standard SomaScan delivery format.
is_lifted()
: Logical. Whether the RFU values in a soma_adat
have been lifted from its original signal space to a new signal space.
The Lin's Concordance Correlation Coefficient (CCC) is calculated
by computing the correlation between post-lift RFU values and the
RFU values generated on the original SomaScan version.
This CCC estimate is a measure of how well an analyte can be bridged
across SomaScan versions.
See vignette("lifting-and-bridging", package = "SomaDataIO")
.
As with the lifting scalars, if you have an annotations file
you may view the analyte-specific CCC values via read_annotations()
.
Alternatively, getSomaScanLiftCCC()
retrieves these values
from an internal object for both "serum"
and "plasma"
.
Newer versions of SomaScan typically have additional content, i.e.
new reagents added to the multi-plex assay that bind to additional proteins.
When lifting to a previous SomaScan version, new reagents that do not
exist in the "earlier" assay version assay are scaled by 1.0, and thus
maintained, unmodified in the returned object. Users may need to drop
these columns in order to combine these data with a previous study
from an earlier SomaScan version, e.g. with collapseAdats()
.
In the inverse scenario, lifting "forward" from a previous, lower-plex
version, there will be extra reference values that are unnecessary
to perform the lift, and a warning is triggered. The resulting data
consists of RFU data in the "new" signal space, but with fewer analytes
than would otherwise be expected (e.g. 11k
space with only 5284
analytes; see example below).
Lin, Lawrence I-Kuei. 1989. A Concordance Correlation Coefficient to Evaluate Reproducibility. Biometrics. 45:255-268.
# `example_data` is SomaScan (V4, 5k)
adat <- head(example_data, 3L)
dim(adat)
getSomaScanVersion(adat)
getSignalSpace(adat)
# perform 'lift'
lift_11k <- lift_adat(adat, "5k_to_11k") # warning
is_lifted(lift_11k)
dim(lift_11k)
# attributes updated to reflect the 'lift'
attr(lift_11k, "Header")$HEADER$SignalSpace
attr(lift_11k, "Header")$HEADER$ProcessSteps