View source: R/readBrukerFlexFile-functions.R
readBrukerFlexFile | R Documentation |
This function reads mass spectrometry data in Bruker Daltonics XMASS format used by Bruker Daltonics mass spectrometer of *flex series (autoflex, microflex, ultraflex).
readBrukerFlexFile(
fidFile,
removeMetaData = FALSE,
useHpc = TRUE,
filterZeroIntensities = FALSE,
keepNegativeIntensities = FALSE,
verbose = FALSE
)
fidFile |
|
removeMetaData |
|
useHpc |
|
filterZeroIntensities |
|
keepNegativeIntensities |
|
verbose |
|
readBrukerFlexFile
has to import the following data to calculating
mass from acqu file:
acqu-value | becomes metaData | description |
$BYTORDA | metaData$byteOrder | endianness of fid file |
$TD | metaData$number | total number of measured time periods |
$DELAY | metaData$timeDelay | first measured intensity after metaData$timeDelay ns |
$DW | metaData$timeDelta | ns between measured time periods |
$ML1 | metaData$calibrationConstants[1] | mass calibration constant |
$ML2 | metaData$calibrationConstants[2] | mass calibration constant |
$ML3 | metaData$calibrationConstants[3] | mass calibration constant |
If High Precision Calibration (HPC) is used, readBrukerFlexFile
needs:
acqu-value | becomes metaData | description |
$HPClBHi | metaData$hpc$limits[“maxMass”] | upper mass threshold |
$HPClBLo | metaData$hpc$limits[“minMass”] | lower mass threshold |
$HPClOrd | metaData$hpc$order | polynomial order |
$HPClUse | metaData$hpc$use | maybe using of HPC? (seems to be always “yes” in our test data) |
$HPCStr | metaData$hpc$coefficients | polynomial coefficients in a string |
readBrukerFlexFile
tries also to import [optional]:
acqu-value | becomes metaData | description |
DATATYPE | metaData$dataType | e.g CONTINUOUS MASS SPECTRUM |
SPECTROMETER/DATASYSTEM | metaData$dataSystem | e.g. Bruker Flex Series |
.SPECTROMETER TYPE | metaData$spectrometerType | e.g. TOF |
.INLET | metaData$inlet | DIRECT |
.IONIZATION MODE | metaData$ionizationMode | e.g. LD+ |
$DATE | metaData$date | same as $AQ_DATE but often only "0" |
$ACQMETH | metaData$acquisitionMethod | path to method file |
$AQ_DATE | metaData$acquisitionDate | acquisition date |
$AQ_mod | metaData$acquisitionMode | acquisition mode |
$AQOP_m | metaData$acquisitionOperatorMode, metaData$tofMode | LINEAR / REFLECTOR |
$ATTEN | metaData$laserAttenuation | laser beam attenuation |
$CMT[1:4] | metaData$comments | comments |
$DEFLON | metaData$deflection | deflection ON/OFF |
$DIGTYP | metaData$digitizerType | type of digitizer |
$DPCAL1 | metaData$deflectionPulserCal1 | deflection pulser cal 1 |
$DPMASS | metaData$deflectionPulserMass | deflection pulser mass |
$FCVer | metaData$flexControlVersion | Version of Bruker Daltonics FlexControl software |
$ID_raw | metaData$id | spectrum id |
$INSTRUM | metaData$instrument | e.g. AUTOFLEX |
$InstrID | metaData$instrumentId | ID of mass spectrometer |
$InstTyp | metaData$instrumentType | instrument type |
$Lift1 | metaData$lift[1] | LIFT constant? |
$Lift2 | metaData$lift[2] | LIFT constant? |
$Masserr | metaData$massError | initial mass error in ppm |
$NoSHOTS | metaData$laserShots | number of applied laser shots |
$PATCHNO | metaData$patch | sample postion on target |
$PATH | metaData$path | original file path (on Bruker *flex series controller PC) |
$REPHZ | metaData$laserRepetition | laser repetition rate in Hz |
$SPOTNO | metaData$spot | same as $PATCHNO
(in older files often empty, that's why readBrukerFlexFile
uses $PATHNO instead) |
$SPType | metaData$spectrumType | e.g. TOF |
$TgIDS | metaData$target$id | target ids |
$TgCount | metaData$target$count | number of measurements with this target |
$TgSer | metaData$target$serialNumber | target serial number |
$TgTyp | metaData$target$typeNumber | target type number |
$TLift | metaData$tlift | LIFT constant? |
import from file path:
value | becomes metaData | description |
full current path to fid file | metaData$file | path on local machine |
sample name | metaData$sampleName | - |
filterZeroIntensities
: Change default value is not recommended!
If TRUE
all intensities equal zero are removed.
This parameter exists only to be compatible to
Bruker Daltonics CompassXport's mzXML export function. For details see:
‘Release Notes for CompassXport 3.0.3’,
cap. 6 ‘Filtering of Zero Intensities’:
“Bruker Daltonics' Acquisition Software will compress Analysis raw
data. To save on operation time and to keep export file sizes small,
CompassXport 3.0.3 will filter out zero (0.0) intensities
when exporting to mzXML or mzData ...”
keepNegativeIntensities
: Change default value is
not recommended! If TRUE
negative intensity values are not
replaced by zero. This parameter exists only to be compatible to
Bruker Daltonics CompassXport.
A list
of spectra and metadata.
spectrum$mass
: A vector of calculated mass.
spectrum$tof
: A vector of time-of-flight data.
spectrum$intensity
: A vector of intensity values.
metaData
: A list of metaData depending on read spectrum.
https://github.com/sgibb/readBrukerFlexData/wiki,
importBrukerFlex
,
readBrukerFlexDir
,
.hpc
## load library
library("readBrukerFlexData")
## get examples directory
exampleDirectory <- system.file("Examples", package="readBrukerFlexData")
## read example spectrum
spec <- readBrukerFlexFile(file.path(exampleDirectory,
"2010_05_19_Gibb_C8_A1/0_A1/1/1SLin/fid"))
## print metaData
print(spec$metaData)
## plot spectrum
plot(spec$spectrum$mass, spec$spectrum$intensity, type="l", col="red")
Add the following code to your website.