title: HGIS status and plan author: Brett Longworth date: 2017-10-27
Tested for flow and discharge with various capillaries. No evidence of discharge with pure CO2. Did not run source with beam. Looks like 2m of 75um capillary should give good flows @ 15kPa (150mbar). 100um capillaries should work well for lower pressures or with Helium dilution. PEEK capillaries seem to produce lower flow rates than they should. May be getting pinched in 1/16 swage fittings.
Ran source with dual inlet. Used 100um PEEK capillary. Delivered gas at pressure range from 0 to 800 mbar. No sparking, best currents (9.5uA) at 600 mbar. Given past performance this confirms that PEEK capillary is getting pinched. Glass capillaries or better PEEK fittings may be needed.
Tested how delivery pressure/flow depletes with bellows and micro-inlet. Used 100um PEEK capillary and leak checker. No measurable depletion over 30min with bellows at 25 mbar. Micro-inlet depletes by about a factor of 3 over 30 min starting at 25 mbar.
Al and Kalina measured volumes in the dual inlet to help quantify ranges of sample size and delivery pressures.
With CO2 pressure set at ~100 mbar...
With the ability to adjust delivery pressure using capillary length, moles of CO2 required can go up or down by an order of magnitude. So:
At 100 mbar:
At 10 mbar
Installed. Need to be tested and code written.
Gate valve installed. Needs control wiring and integration. Need to install pump and test.
Next test should probably be more ion source tests with low pressure capillaries. We should test delivery at 10 mbar and take live/dead gas data for various configurations.
On the Helium front: install turbopump, test operation. Initial tests of He mixing and delivery.
Add the following code to your website.
For more information on customizing the embed code, read Embedding Snippets.