| ph_calibration | R Documentation | 
Fit pH calibration parameters and generate new configuration file
This function estimates pH calibration parameters using Sea-Bird's calibration equation.
ph_calibration(
  ph,
  temperature,
  ph_voltage,
  xmlcon_in,
  xmlcon_out = NULL,
  ph_serial_number
)
ph | 
 ph of the reference standard  | 
temperature | 
 Reference standard temperature in degrees celcisus  | 
ph_voltage | 
 Measured pH channel voltage at the reference standard  | 
xmlcon_in | 
 Filepath to the current CTD configuration file.  | 
xmlcon_out | 
 Filepath to the output CTD configuration file.  | 
ph_serial_number | 
 Serial number for the pH sensor (typically four digits.)  | 
Sean Rohan
# Example of pH calibration with a demo file
ph_calibration(xmlcon_in = system.file("./extdata/example/SBE19plusV2_8106_ph_DO_leg2.xmlcon", package = "gapctd"),
               ph_serial_number = 9090,
               ph_voltage = c(1.977, 2.797, 3.534),
               ph = c(4, 7, 10),
               temperature = 23)