centWave Parameters

ppm, peakwidth, snthresh, prefilter, mzCenterFun, integrate, mzdiff, fitgauss, scanrange, noise.

 

All centWave parameters require fine-tuning for each assay type and mass analyser. The following provides a short description of each parameter value and it's relevance in the centWave algorithm.

 

| Parameter | Description | Pre-adjusted value | |-----------|-------------|:--------:| |ppm| Allowed signal deviation in m/z dimension | 25 | |peakwidth | Range of peak elution times (in seconds) | 20-50 s | |snthresh | Threshold signal to noise ratio | 10 | |prefilter| Peak definition: Number of data points (n) exceeding a certain intensity threshold (I)| k=3, I=100 | |mzdiff| Accepted closeness of two signals in m/z dimension |-0.001 | |noise| Intensity cut-off, values below are considered as instrument noise | 0 | |mzCenterFun| Function to calculate the m/z center of a chromatographic peak | weighted Mean (wMean) | |integrate| Integration method for peak quantification: 1 - Mexican hat filtered data, 2 - real data | 1 | |fitgauss| Peak parameterisation using Gaussian distribution | FALSE | |scanrange| Perform peak picking in a scan range interval | numeric(0) |

 

 

 

ppm: Maximum allowed signal deviation in m/z dimension

This centWave parameter ppm specifies the tolerance in m/z values for defining a signal in m/z dimensions. This parameter is closely related to the mass accuracy of the mass spectrometer, which is traditionally expressed in parts per million (ppm).

![](man/figures/pars1/ppm_formula.png)

Higher ppm values allow more m/z value variations and lead to detection of a higher number of potential peaks - increasing false positives. Lower values can lead to missing out on many signals, since the measured mass values differ from the true mass more than expected.

The illustration below shows a peak picking example using the same LC-MS data, where ppm parameter values was changed and all other centWave parameters were held constant.

![](man/figures/pars1/ppm_comp.png)

 

 

 

mzdiff: Accepted closeness of two signals in m/z dimension

The mzdiff parameter specifies the allowed minimum distance of two co-eluting peaks in m/z dimension. An mzdiff value of 1 indicates that the m/z value of two signals with overalapping scan times be at least 1, in order for both signals to be included in the resulting peak list.

The centWave mzdiff parameter can also take negative values, indicating that the same data point can be allocated to two different peaks. Assigning negative mzdiff values has implications for further downstream processing steps, e.g., establishing correspondence of overlapping peaks across different samples. Higher mzdiff parameter values lead to detecting more features (potential signals), lower values reduce the number of features detected.

Below is an example using the same data processed with mzdiff values of 1 and 0.01.

![](man/figures/pars1/mzdiff.png)

 

 

 

noise: Intensity cut-off, values below are not considered

The ion detection in most mass spectrometers is accomplished with electron multipliers. These instruments are highly sensitive and produce electronical noise, which is visible as (usually) random data points below a certain intensity cut-off.

The noise structure of LC-MS spectra generated with mass specs of different types and from different vendors (incl. software updates) can be inherently different. Therefore, this parameter requires careful adjustment for each mass spectrometer setup. Lower values increases centWave computation time, higher values lead to missing out on true data signals.

![](man/figures/pars1/noise.png)

 

 

 

snthresh: Threshold of signal to noise ratio

Closely related to the noise structure is the snthresh parameter, allowing to set a minimal signal to noise ratio (S/N) for peaks to be detected. snthresh builds on an S/N estimate that defines noise intensities locally in the scan time dimension:

![](man/figures/pars1/snthresh_formula.png)

This S/N definition can be problematic when the noise structure is not evenly distributed across the m/z dimension. The example below shows that higher intensity signals are discarded with higher snthresh values, most likely due to noisy and high intensity data points in proximity to the signal's m/z trace. The final peak list using a high snthresh parameter value contains signals of low intensities, which is somewhat unexpected. UPDATE: THIS BEHAVIOUR IS ONLY OBSERVERVED WHEN CENTWAVE FUNCITON IS PARAMTETRISED WITH THE SCANRANGE ARGUMENT.

![](man/figures/pars1/snthresh_comp.png)

 

 

 

peakwidth: Range of peak elution times

The peakwidth parameter allows specification of the minimal and maximal peak elution time (in seconds). Setting the minimal elution time to higher values risks missing out small signals (usually of low intensity). However, the literature suggest a minimum number of six data points in order to quantify a peak reliably. Therefore, the optimal minimal peakwidth parameter is related to the number of scans on MS 1 level (this is a user-defined setting of the mass spectrometer).

The example below shows the results of centWave peak picking performed with different peakwidth parameter values and where all other parameters were held constant.

An unexpected algorithm behaviour was observed when setting the minimal elution time to a value of 1, which resulted in peak splitting of a coherent signal which was not split with higher peakwidth values.

![](man/figures/pars1/peakwidth.png)

 

 

 

prefilter: Number of data points (k) exceeding a certain intensity threshold (I)

The prefilter parameter is similar to the noise parameter, but instead of specifying an intensity cut-off, it specifies a minimal number of data points (k) that exceed a certain intensity value (I). A signal is discarded if it is represented by less than k consecutive data points of intensity I.

Lower prefilter parameter values perform less filtering of signals with low intensity / scan time, and therefore, reduce computation time. The values of this parameter strongly depend on the characteristics of well-behavied LC-MS signals. This parameter should not be set to stringent as it can lead to discarding true positive signals.

![](man/figures/pars1/prefilter.png)

 

 

 

mzCenterFun: m/z summary statistic of a peak

In the final LC-MS feature table, each feature is characterised by a scan time and m/z value. The latter represents a summary statistic of all data points defining a feature, where each one has a slightly different m/z value (see also ppm parameter). The mzCenterFun parameter specifies the m/z summary statistic of a feature. In practice, the different mzCenterFun options have very little impact.

![](man/figures/pars1/mzCenterFun.png)

 

 

 

integrate: Integration method for peak quantification

Signal quantification is performed by integrating the area under a peak. The integrate function specifies if the non-transformed data should be used for peak integration or if a wavelet-transform filtered data should be used (upper and lower panel, respectively, in the plot below). The former is sensitive to outliers and noise, since centWave defines left and right peak boundaries through change in slope (see Figure below). The latter is less sensitive to noise, however, the wavelet-filtered data is less exact and can lead to signal mis-representations.

![](man/figures/pars1/integrate.png)

 

 

 

scanrange: Perform peak picking in a specific scan range

The scanrange parameter allows to select a scan time interval for peak picking of the entire m/z range. Spectral data outside the specified interval are not considered for peak picking. This parameter is usually not specified in untargeted metabolic profiling studies. The scanrange parameter takes scan ID values (it is not specified in seconds).

![](man/figures/pars1/scanrange.png)