Description Usage Arguments Details Value
Fits an ITC experiment to a Wiseman isotherm(1) to determine thermodynamic statistics. Raw ".itc" formatted data should be read in using "read.itc". Fits the data to a thermodynamic model then prints a graphical summary of the fit.
1 2 3 4 5 6 7 8 9 | MetaboMgITC(
cell,
blank,
Thermodynamic.equation = "Wiseman.isotherm",
Fit.start = list(H = 2.534, K = 2880),
Remove.injection = 1,
Saturation.threshold = FALSE,
Save.path.and.prefix = FALSE
)
|
cell |
Object created by "read.itc" for the ITC run containing the macromolecule. |
blank |
Object created by "read.itc" for the ITC run containing the ligand tirated into buffer. |
Thermodynamic.equation |
The thermodynamic model that you want to fit the data to. Default = "wiseman.isotherm". |
Fit.start |
Starting parameters for the non-linear regression. Default = = list(H = 2.534, K = 2880). |
Remove.injection |
Injections you want to remove from subsequent analysis. Default = 1 is the standard for ITC experiments. More than one injection can be supplied in a vector. |
Saturation.threshold |
For low c value ITC experiments fit to Wiseman isotherms, macomolecule saturation ranging from 70% to 90% produce the same answers. Thus, sometimes it is desirable to standardize the saturation threshold between experiments to minimize degredation of a ligand or macromolecule (for example with ATP binding Mg). Default = FALSE. If set to 0.8, this function will only fit data required to reach 80% macromolecule saturation. |
(1) Turnbull, W. B.; Daranas, A. H. On the Value of c: Can Low Affinity Systems Be Studied by Isothermal Titration Calorimetry? Journal of the American Chemical Society 2003, 125 (48), 14859–14866. https://doi.org/10.1021/ja036166s.
A list containing summary statistics for the fit and a graphical depiction of the fits, the raw ITC curve, and the summary statistics.