| nacho_thresholds | R Documentation |
Builds the thresholds that load_rcc() and normalise() use to flag
samples.
nacho_thresholds(
instrument = c("max", "flex", "pro", "sprint"),
preset = c("nsolver", "legacy"),
haemolysis = FALSE
)
instrument |
The nCounter instrument: |
preset |
|
haemolysis |
If |
"nsolver" follows the current Bruker nCounter knowledge base
(https://brukerspatialbiology.com/support/, "nCounter QC flags"): binding density
0.05 to 2.25 on MAX, FLEX and PRO and 0.1 to 1.8 on SPRINT, field of view at
least 75 %, positive control linearity at least 0.95 (POS_A to POS_E, with
log2(count + 1)), limit of detection at least 2, positive factor 0.3 to 3
and housekeeping factor 0.1 to 10.
It also asks for at least three housekeeping genes above background, as
Bruker asks (Housekeeping_detected).
"legacy" reproduces NACHO 2: binding density 0.1 to 2.25 on every
instrument, positive control linearity with POS_F, positive factor 1/4 to 4
and housekeeping factor 1/11 to 11.
With background = "geo", background_mode = "subtract", it gives back the
NACHO 2 outlier calls.
The preset also chooses how negative probes are excluded and how the positive control linearity is computed, so changing it normalises again.
For miRNA panels, Ligation_order (the three ligation positive controls
in decreasing order), Ligation_R2 (their log2 counts on a line) and
Ligation_NEG (the largest ligation negative minus the detection limit)
are NACHO's own definitions, since Bruker publishes no threshold.
"legacy" never flags them.
Whole-blood RNA raises LIG_NEG_A, and to a lesser extent LIG_NEG_B,
together with red blood cell miRNAs such as miR-451a.
The ligation negatives of such samples are above the detection limit, so
most samples of a whole-blood study fail Ligation_NEG.
This pattern occurs in independent PAXgene whole-blood studies on GEO
(GSE303669 and GSE317496), and not in plasma, serum or extracellular
vesicle studies.
When more than half of the samples fail Ligation_NEG, load_rcc()
warns about this pattern.
For whole-blood samples, set a higher upper bound for Ligation_NEG and
pass the thresholds to normalise(outliers_thresholds = ).
Do not remove these samples with exclude_outliers() because of this
flag only.
Each element can be changed afterwards; two numbers are a lower and an
upper bound, and one number is a lower bound. Use -Inf or Inf for an
open bound.
A named list: preset, instrument, then one element per metric.
nacho_thresholds()
nacho_thresholds("sprint")
nacho_thresholds(preset = "legacy")
Add the following code to your website.