Fits a univariate calibration curve (signal against concentration), with its figures of merit, limits of detection and quantification, and tests of linearity. The curve predicts the concentration of new samples from their signal, with confidence intervals.
calibration_curve(
data,
signal,
concentration,
model = "linear",
weights = NULL,
blank = NULL,
lod_method = NULL,
level = 0.95,
regression = "ols",
concentration_se = NULL,
signal_se = NULL,
variance_ratio = NULL,
lod_convention = "ich",
alpha = 0.05,
beta = 0.05,
replicates = 1
)
data |
A data frame with the calibration samples. |
signal, concentration |
The columns of |
model |
|
weights |
|
blank |
An optional numeric vector of replicate signals of a blank sample. |
lod_method |
The estimate of the blank standard deviation:
|
level |
The confidence level of the intervals of the coefficients. Default is 0.95. |
regression |
|
concentration_se, signal_se |
For |
variance_ratio |
For |
lod_convention |
The convention of the limits of detection and
quantification: |
alpha, beta |
For |
replicates |
For |
The signal, such as a line intensity from line_intensities() or a line
ratio, is modeled as a straight line or a second-degree polynomial of the
concentration, fitted by (weighted) least squares. Weights "1/x" or
"1/x2" suit signals whose variance grows with concentration, which is
common in LIBS.
Figures of merit. The sensitivity is the slope of the curve at zero
concentration. The limits of detection and quantification are
LOD = 3.3\,\sigma / S and LOQ = 10\,\sigma / S (ICH, 2005),
where S is the sensitivity and \sigma the standard deviation
of the signal of a blank, estimated from
"residual": the residual standard deviation of the fit (default
without blank);
"intercept": the standard error of the intercept;
"blank": the standard deviation of replicate blank signals
(default when given).
With weights, the residual standard deviation depends on the scale of the
weights, so prefer "blank" or "intercept".
With lod_convention = "iso11843", the limits follow ISO 11843-2 and
IUPAC (Currie, 1999), and account for the uncertainty of the curve and
for the false positive and false negative risks \alpha and
\beta. The standard deviation of a net signal near zero is
\sigma_0 = \sqrt{\sigma^2 / K + s_a^2}, where \sigma is the
standard deviation of the signal ("residual" or "blank"), K the
number of replicates averaged for a sample and s_a the standard
error of the intercept. The critical value (decision limit), above which
an element is detected, is x_C = t_{1-\alpha,\nu}\,\sigma_0 / S; the
limit of detection, the content detected with probability 1 -
\beta, is x_D = \delta_{\nu,\alpha,\beta}\,\sigma_0 / S, where
\delta is the noncentrality parameter of the noncentral t
distribution (about 2 t_{1-\alpha,\nu} for \alpha = \beta) and
\nu the degrees of freedom of \sigma; and the limit of
quantification (IUPAC) is x_Q = 10\,\sigma_0 / S. For a weighted
curve, the standard deviation must come from blank signals.
Errors in the concentrations. Least squares assume exact
concentrations. When those of the standards are uncertain, such as the
certified values of reference materials, the slope is biased toward zero
and the intervals too narrow. regression = "york" fits the line by the
method of York et al. (2004), with the standard uncertainties of the
concentration (concentration_se) and of the signal (signal_se) of
each standard; its standard errors are multiplied by the square root of
the reduced chi-squared when it exceeds 1 (the points scatter more than
their uncertainties explain). regression = "deming" (Deming
regression) assumes constant uncertainties, whose ratio
variance_ratio (variance of the signal over variance of the
concentration, in their units) is given, or computed from the mean
squared signal_se and concentration_se; its standard errors come from
the scatter of the points. Both fit straight lines without weights; the
tests of linearity are those of the least-squares fit.
Linearity. For a straight line, Mandel's test compares its residual variance with that of a quadratic fit; a small p-value means that the curvature is significant (as with self-absorption or detector saturation). When concentrations are replicated, the lack-of-fit test compares the residuals with the pure error of the replicates.
Intervals. The confidence band of the curve shows where the mean
signal lies at each concentration; the prediction band, wider, where a
single new measurement is expected to fall, since it adds the noise of
the measurement. plot_calibration() draws either or both, and
predict() computes them for given
concentrations (type = "signal").
Inverse prediction. The concentration of a sample is the solution of
the calibration equation for its signal (for a quadratic curve, the root
within or nearest to the calibration range). Its standard error, by the
delta method, combines the uncertainty of the curve with that of the new
signal, the mean of replicates measurements.
An object of class specproc_calibration, a list with
coefficients: a tibble of the estimates, standard errors and
confidence intervals (lower, upper) at level;
figures_of_merit: a tibble with the number of standards n, the
sensitivity, r_squared, the residual standard deviation sigma,
the blank standard deviation used (sigma_blank), the
critical_value (ISO 11843 only, NA otherwise), lod and loq (in
concentration units);
linearity: a tibble with the statistic, degrees of freedom and
p-value of Mandel's test and of the lack-of-fit test (when they apply);
fit: the least-squares lm fit, eiv, the Deming or York fit
(NULL for least squares), with its chi_squared, and data, the
calibration data.
Use predict() for new samples and
plot_calibration() to draw the curve.
Christian L. Goueguel
Currie, L.A. (1999). Detection and quantification limits: origins and historical overview. Analytica Chimica Acta, 391(2):127-134.
ICH (2005). Validation of analytical procedures: text and methodology Q2(R1). International Conference on Harmonisation.
ISO 11843-2 (2000). Capability of detection, Part 2: Methodology in the linear calibration case. International Organization for Standardization, Geneva.
Mandel, J. (1964). The Statistical Analysis of Experimental Data. Interscience, New York.
Miller, J.N., Miller, J.C. (2018). Statistics and Chemometrics for Analytical Chemistry, 7th ed. Pearson, Harlow.
York, D., Evensen, N.M., López Martínez, M., De Basabe Delgado, J. (2004). Unified equations for the slope, intercept, and standard errors of the best straight line. American Journal of Physics, 72(3):367-375.
predict.specproc_calibration(), plot_calibration(),
line_intensities(), nas()
data(forageLIBS)
# the K I 769.90 nm line, normalized to the C I 247.86 nm line of the matrix
lines <- line_intensities(forageLIBS[-(1:14)], c(C = 247.856, K = 769.896), baseline = TRUE)
standards <- data.frame(
K = forageLIBS$K,
signal = lines$intensity[lines$line == "K"] / lines$intensity[lines$line == "C"]
)
cal <- calibration_curve(standards[1:300, ], signal, K)
cal
# the concentrations of new samples, from their signals
head(predict(cal, standards$signal[301:368]))
# limits of detection by ISO 11843-2
calibration_curve(standards[1:300, ], signal, K, lod_convention = "iso11843")
# reference materials with certified values known within 3%, and
# signals within 2% (simulated)
set.seed(1)
crm <- data.frame(certified = c(0.5, 1, 2, 4, 6, 8))
crm$signal <- 0.05 + 0.2 * crm$certified * (1 + rnorm(6, sd = 0.02)) +
0.2 * rnorm(6, sd = 0.03 * crm$certified)
crm$certified_se <- 0.03 * crm$certified
crm$signal_se <- 0.02 * crm$signal
calibration_curve(crm, signal, certified, regression = "york",
concentration_se = certified_se, signal_se = signal_se)
Add the following code to your website.
For more information on customizing the embed code, read Embedding Snippets.