View source: R/correct_self_absorption.R
| correct_self_absorption | R Documentation |
Corrects the intensities of emission lines for self-absorption by the internal reference method of Sun and Yu (2009): for each species, the self-absorption coefficient of every line is measured against a line of the same species that is little self-absorbed.
correct_self_absorption(
lines,
temperature = NULL,
electron_density = NULL,
ionization_energy = NULL,
reference = NULL,
units = "energy",
tol = 1e-08,
timeout = 120
)
lines |
A data frame with one row per line and columns |
temperature |
The plasma temperature, in K. Without it, it is
estimated from the corrected lines, which needs |
electron_density |
The electron density, in cm |
ionization_energy |
The ionization energies of the neutral atoms,
in eV, named by element, to estimate the temperature. By default, they
are taken from the NIST database (see |
reference |
The reference lines: row numbers of |
units |
The units of |
tol |
The relative tolerance on the estimated temperature. Default is 1e-8. |
timeout |
The timeout of the NIST downloads, in seconds. |
In an optically thin plasma in LTE, the intensity ratio of two lines of
the same species depends only on their atomic data and on the
temperature. With a reference line r assumed optically thin, the
self-absorption coefficient of a line j is
SA_j = \frac{I_j}{I_r} \frac{g_r A_r \lambda_j}{g_j A_j \lambda_r}
\exp\left(\frac{E_j - E_r}{k_B T}\right)
(for intensities in energy units; without the wavelengths for photon
counts), where g, A and E are the statistical weight,
transition probability and energy of the upper level. The corrected
intensity is I_j / SA_j.
Self-absorption grows with the optical depth of the line, roughly
proportional to g_k A_{ki} \lambda^4 \exp(-E_i / k_B T), where
E_i is the energy of the lower level: resonance lines (E_i =
0) and strong lines are the most absorbed. By default, the reference of
each species is the line of smallest optical depth, which needs the
lower-level energies (column Ei, as returned by nist_lines());
otherwise, give the reference lines with reference. The reference line
must be well measured: its noise propagates to every coefficient.
The corrected lines of a species lie, by construction, on the Boltzmann
plot of its reference line at the temperature used, so they cannot give
the temperature by themselves. Give temperature, measured on optically
thin lines (for example with saha_boltzmann() or cf_libs()), or
give electron_density: the temperature is then the one at which the
corrected lines of the neutral atom and of the ion of each element lie on
a common Saha-Boltzmann plot (as in cf_libs()), which needs lines of
both stages for at least one element. This is the iterative scheme of Sun
and Yu (2009), solved directly for the temperature.
Coefficients above 1 mean that the line is stronger than expected from the reference, because of noise, errors of the atomic data, or a self-absorbed reference; they are kept, with a warning.
lines, as a tibble, with the corrected intensity, the
measured_intensity, the self-absorption coefficient SA and a
logical reference, ready for boltzmann() or cf_libs(). The
temperature used is in the attribute temperature. A column
intensity_se (as returned by line_intensities()) becomes the
standard error of the corrected intensity, which is computed from the
reference line alone: its relative standard error is that of the
reference line.
Christian L. Goueguel
Sun, L., Yu, H. (2009). Correction of self-absorption effect in calibration-free laser-induced breakdown spectroscopy by an internal reference method. Talanta, 79(2):388-395.
self_absorption() for the coefficient from line widths,
cf_libs(), line_intensities()
data(forageLIBS)
mean_spectrum <- colMeans(forageLIBS[-(1:14)])
# potassium lines: the resonance doublet at 766.49 and 769.90 nm ends on
# the ground state (Ei = 0) and is strongly reabsorbed
k_lines <- data.frame(
species = "K I", wavelength = c(404.414, 404.721, 691.108, 693.877, 766.490, 769.896),
Aki = c(1.150e6, 1.070e6, 2.500e6, 4.956e6, 3.779e7, 3.734e7),
gk = c(4, 2, 2, 2, 4, 2),
Ek = c(3.065, 3.063, 3.403, 3.403, 1.617, 1.610),
Ei = c(0, 0, 1.610, 1.617, 0, 0)
)
corrected <- correct_self_absorption(line_intensities(mean_spectrum, k_lines), temperature = 8000)
corrected[c("wavelength", "measured_intensity", "SA", "reference")]