Transfers spectra from a secondary instrument to the primary instrument, on which the calibration model was built, by direct standardization (DS; Wang et al., 1991): a linear map of the whole secondary spectrum onto the whole primary spectrum, estimated on transfer samples measured on both instruments.
ds(x1, x2, ncomp = NULL, tol = 1e-08)
x1 |
The spectra of the transfer samples on the primary instrument: a numeric matrix or data frame, one spectrum per row. |
x2 |
The spectra of the same samples, in the same order, on the secondary instrument. |
ncomp |
The number of singular vectors of the secondary spectra kept. By
default, all those with a singular value above |
tol |
The relative tolerance on the singular values. Default is 1e-8. |
With the transfer spectra centered by the mean of each instrument,
\textbf{X}_1 \approx \textbf{X}_2 \textbf{F}, and the transfer
matrix is \textbf{F} = \textbf{X}_2^{+} \textbf{X}_1, where
\textbf{X}_2^{+} is the pseudo-inverse of the secondary spectra,
truncated to their first ncomp singular vectors. A new secondary spectrum
is transferred as \bar{x}_1 + (x_2 - \bar{x}_2) \textbf{F}.
With as many components as the transfer samples allow (ncomp = NULL),
the transfer spectra are reproduced exactly, and the noise of the secondary
spectra is amplified: fewer components give a smoother, more robust
transfer. Unlike pds(), DS mixes all wavelengths, so it corrects global
differences of the response but needs more transfer samples. The transfer
matrix (p \times p) is kept in factored form, as ncomp columns of
each factor.
An object of class specproc_transfer. Use
predict() to transfer secondary spectra.
Christian L. Goueguel
Wang, Y., Veltkamp, D.J., Kowalski, B.R. (1991). Multivariate instrument standardization. Analytical Chemistry, 63(23):2750-2756.
pds(), sst(), slope_bias(), predict.specproc_transfer()
data(forageLIBS)
wl <- suppressWarnings(as.numeric(names(forageLIBS)))
spectra <- as.matrix(forageLIBS[which(wl > 760 & wl < 780)]) # the K I lines
# a second instrument, with a shifted and less sensitive response
secondary <- 0.8 * cbind(spectra[, -1], spectra[, ncol(spectra)]) + 50
transfer <- ds(spectra[1:30, ], secondary[1:30, ], ncomp = 10)
transfer
# the other samples, measured on the secondary, transferred to the primary
transferred <- predict(transfer, secondary[31:368, ])
mean(abs(transferred - spectra[31:368, ])) / mean(spectra)