correct_response: Correct Spectra or Line Intensities for the Spectral Response

View source: R/response.R

correct_responseR Documentation

Correct Spectra or Line Intensities for the Spectral Response

Description

Divides spectra, or the intensities of emission lines, by the relative spectral response of the instrument measured by response_calibration().

Usage

correct_response(x, response)

Arguments

x

Spectra (a numeric vector named by wavelength, or a matrix or data frame with one spectrum per row and the wavelengths as column names), or a table of lines with wavelength and intensity columns.

response

The result of response_calibration().

Details

Spectra must have the wavelength axis of the response (the same channels, in the same order): each channel is divided by its response.

A table of lines, such as returned by line_intensities(), is recognized by its wavelength and intensity columns. The intensity (and height) of each line is divided by the response at the channel of its peak (peak_wavelength, or else the channel nearest to its wavelength), and its standard error intensity_se (created if absent) combines in quadrature its relative standard error and that of the response. Correcting the line intensities rather than the spectra keeps the noise estimates of line_intensities() valid. The errors of the response of nearby lines are correlated, which plasma_uncertainty() does not account for.

Correct the intensities for the response before correct_self_absorption(), boltzmann() or cf_libs().

Value

x, corrected. For a table of lines, with the corrected intensity (and height), its standard error intensity_se, and the response used. Channels or lines without response become NA.

See Also

response_calibration(), line_intensities()

Examples

planck <- function(wl, t) 1 / (wl^5 * (exp(1.4388e7 / (wl * t)) - 1))
reference <- data.frame(wavelength = seq(250, 850, by = 10))
reference$value <- planck(reference$wavelength, 3100)
reference$rsd <- 0.03
wl <- seq(260, 840, by = 0.1)
true_response <- exp(-((wl - 500) / 250)^2)
set.seed(1)
lamp <- 2e4 * true_response * planck(wl, 3100) / planck(500, 3100) + rnorm(length(wl), sd = 20)
names(lamp) <- wl
response <- response_calibration(lamp, reference)

# a spectrum with two lines of equal emission, at 300 and 650 nm
emission <- 1000 * exp(-(wl - 300)^2 / 0.01) + 1000 * exp(-(wl - 650)^2 / 0.01)
spectrum <- stats::setNames(true_response * emission + rnorm(length(wl)), wl)
lines <- line_intensities(spectrum, c(a = 300, b = 650))
lines$intensity   # unequal: the response differs
corrected <- correct_response(lines, response)
corrected[c("line", "intensity", "intensity_se", "response")]

# or the whole spectrum
corrected_spectrum <- correct_response(spectrum, response)