#' @title Spectral Response from a Calibration Lamp
#'
#' @author Christian L. Goueguel
#'
#' @description
#' Measures the relative spectral response of an instrument (the
#' efficiency of the collection optics, fiber, grating and detector at each
#' wavelength) from spectra of a calibration lamp of known spectral
#' irradiance or radiance, such as a deuterium-halogen lamp, with its
#' uncertainty. [correct_response()] then corrects spectra or line
#' intensities.
#'
#' @details
#' The response at each channel is the ratio of the measured lamp spectrum
#' (after subtraction of `dark`) to the certified irradiance of the lamp,
#' interpolated at the channel. The certified values, often tabulated every
#' few nanometers, are interpolated by a monotone cubic spline of their
#' logarithm; channels outside their range get no response. The ratio is
#' smoothed by a local linear fit over `smooth` nm, within each detector
#' segment when `segments = TRUE` (the axis is split where the wavelengths
#' step back or jump, as in [wavelength_calibration()]), so that the noise
#' of the lamp spectrum is averaged without biasing the steep edges of the
#' response. The channels in the `exclude` ranges, such as the Balmer lines
#' of a deuterium lamp (486.0 and 656.1 nm) that the certificate leaves
#' out, are interpolated from their neighbors. The response is scaled to a
#' maximum of 1.
#'
#' The relative standard uncertainty of the response (`rsd`) combines in
#' quadrature that of the certified values (column `rsd` of `reference`,
#' interpolated linearly) and that of the measurement: with several lamp
#' spectra, the standard error of their mean response (each spectrum
#' scaled to a common level first, so that drifts of the lamp output do not
#' count); with one, the noise of the spectrum (estimated from its
#' channel-to-channel differences) averaged over the smoothing window.
#' Certificates often give expanded uncertainties (coverage factor 2):
#' divide them by 2.
#'
#' The lamp must be measured through the same optical path as the plasma
#' (collection optics, fiber, spectrometer settings), and its spectrum
#' must not be saturated. With a certificate in energy units (W per nm),
#' the corrected intensities are proportional to the spectral radiance
#' (energy units, as `units = "energy"` of [boltzmann()]).
#'
#' @param lamp The spectra of the calibration lamp: a numeric vector named
#' by wavelength, or a matrix or data frame with one spectrum per row and
#' the wavelengths as column names (other columns are ignored). Several
#' spectra are averaged.
#' @param reference The certified spectral irradiance (or radiance) of the
#' lamp: a data frame with columns `wavelength` (nm) and `value`, and
#' optionally `rsd`, the relative standard uncertainties of the values
#' (0.03 for 3%).
#' @param dark An optional dark spectrum (or several, averaged) on the same
#' axis as `lamp`, subtracted from it.
#' @param smooth The width of the smoothing window, in nm. Default is 1; 0
#' turns the smoothing off.
#' @param exclude Wavelength ranges left out of the response and
#' interpolated: a numeric vector of length 2, a list of them, or a
#' two-column matrix. Default is `NULL`.
#' @param segments A logical: smooth each detector segment separately
#' (`TRUE`, default).
#'
#' @return An object of class `specproc_response`, a list with
#' - `response`: a tibble with, for each channel in column order, its
#' `wavelength`, detector `segment`, relative `response` (`NA` without
#' response), and its relative standard uncertainty `rsd`, with the
#' parts from the reference (`reference_rsd`) and from the measurement
#' (`measurement_rsd`);
#' - `n_spectra`, `smooth`, `exclude` and the wavelength `range` of the
#' reference.
#'
#' Use [correct_response()] to correct spectra or line intensities, and
#' [plot_response()] to draw the response.
#'
#' @seealso [correct_response()], [plot_response()], [line_intensities()]
#' @export response_calibration
#'
#' @examples
#' # the certified spectral irradiance of a lamp, every 10 nm, known within
#' # 3% (simulated here as a blackbody at 3100 K)
#' planck <- function(wl, t) 1 / (wl^5 * (exp(1.4388e7 / (wl * t)) - 1))
#' reference <- data.frame(wavelength = seq(250, 850, by = 10))
#' reference$value <- planck(reference$wavelength, 3100)
#' reference$rsd <- 0.03
#'
#' # five spectra of the lamp, through an instrument of unknown response
#' wl <- seq(260, 840, by = 0.1)
#' true_response <- exp(-((wl - 500) / 250)^2)
#' set.seed(1)
#' lamp <- t(replicate(5, 2e4 * true_response * planck(wl, 3100) / planck(500, 3100) +
#' rnorm(length(wl), sd = 20)))
#' colnames(lamp) <- wl
#'
#' response <- response_calibration(lamp, reference)
#' response
#' plot_response(response)
response_calibration <- function(lamp, reference, dark = NULL, smooth = 1, exclude = NULL,
segments = TRUE) {
sp <- line_spectra(lamp)
x <- sp$x
wl <- sp$wavelength
if (!is.null(dark)) {
dk <- line_spectra(dark)
if (length(dk$wavelength) != length(wl) || max(abs(dk$wavelength - wl)) > 1e-6) {
stop("'dark' must have the wavelengths of 'lamp'.", call. = FALSE)
}
x <- sweep(x, 2, colMeans(dk$x, na.rm = TRUE))
}
reference <- check_response_reference(reference)
check_number(smooth, "smooth", lower = 0)
check_flag(segments, "segments")
ranges <- exclude_ranges(exclude)
excluded <- vapply(wl, function(w) any(w >= ranges[, 1] & w <= ranges[, 2]), logical(1))
value <- reference_values(reference, wl)
reference_rsd <- if ("rsd" %in% names(reference)) {
stats::approx(reference$wavelength, reference$rsd, xout = wl)$y
} else {
ifelse(is.na(value), NA_real_, 0)
}
outside <- is.na(value)
if (any(outside)) {
warning(sum(outside), " channel(s) outside the range of the reference (",
paste(format(range(reference$wavelength)), collapse = " to "),
" nm) have no response.", call. = FALSE)
}
# ratio of each lamp spectrum to the reference, scaled to a common level
ratio <- sweep(x, 2, value, "/")
ratio[, excluded | outside] <- NA
ratio[!is.finite(ratio)] <- NA
level <- apply(ratio, 1, stats::median, na.rm = TRUE)
if (any(!is.finite(level) | level <= 0)) {
stop("A lamp spectrum has no positive signal within the range of the reference.",
call. = FALSE)
}
ratio <- ratio / level
seg <- if (segments) wavelength_segments(wl) else rep(1L, length(wl))
fits <- lapply(seq_len(nrow(ratio)), function(i) smooth_response(ratio[i, ], wl, seg, smooth))
smoothed <- do.call(rbind, lapply(fits, `[[`, "fit"))
n <- nrow(smoothed)
mean_ratio <- colMeans(smoothed)
measurement_rsd <- if (n > 1) {
apply(smoothed, 2, stats::sd) / sqrt(n) / mean_ratio
} else {
# noise of the lamp spectrum, averaged over the channels of the window
noise <- spectrum_noise(x[1, ])
rel <- noise / (mean_ratio * level * value) / sqrt(fits[[1]]$n)
rel[fits[[1]]$n == 0] <- NA
fill_segments(rel, wl, seg)
}
mean_ratio[outside] <- NA
measurement_rsd[outside] <- NA
mean_ratio[!is.na(mean_ratio) & mean_ratio <= 0] <- NA
response <- mean_ratio / max(mean_ratio, na.rm = TRUE)
measurement_rsd[is.na(response)] <- NA
reference_rsd[is.na(response)] <- NA
response_rsd <- sqrt(reference_rsd^2 + measurement_rsd^2)
structure(
list(response = tibble::tibble(wavelength = wl, segment = seg, response = response,
rsd = response_rsd, reference_rsd = reference_rsd,
measurement_rsd = measurement_rsd),
n_spectra = n, smooth = smooth, exclude = ranges,
range = range(reference$wavelength)),
class = "specproc_response"
)
}
#' @title Correct Spectra or Line Intensities for the Spectral Response
#'
#' @description
#' Divides spectra, or the intensities of emission lines, by the relative
#' spectral response of the instrument measured by
#' [response_calibration()].
#'
#' @details
#' Spectra must have the wavelength axis of the response (the same
#' channels, in the same order): each channel is divided by its response.
#'
#' A table of lines, such as returned by [line_intensities()], is
#' recognized by its `wavelength` and `intensity` columns. The `intensity`
#' (and `height`) of each line is divided by the response at the channel of
#' its peak (`peak_wavelength`, or else the channel nearest to its
#' `wavelength`), and its standard error `intensity_se` (created if
#' absent) combines in quadrature its relative standard error and that of
#' the response. Correcting the line intensities rather than the spectra
#' keeps the noise estimates of [line_intensities()] valid. The errors of
#' the response of nearby lines are correlated, which [plasma_uncertainty()]
#' does not account for.
#'
#' Correct the intensities for the response before
#' [correct_self_absorption()], [boltzmann()] or [cf_libs()].
#'
#' @param x Spectra (a numeric vector named by wavelength, or a matrix or
#' data frame with one spectrum per row and the wavelengths as column
#' names), or a table of lines with `wavelength` and `intensity` columns.
#' @param response The result of [response_calibration()].
#'
#' @return `x`, corrected. For a table of lines, with the corrected
#' `intensity` (and `height`), its standard error `intensity_se`, and the
#' `response` used. Channels or lines without response become `NA`.
#'
#' @seealso [response_calibration()], [line_intensities()]
#' @export correct_response
#'
#' @examples
#' planck <- function(wl, t) 1 / (wl^5 * (exp(1.4388e7 / (wl * t)) - 1))
#' reference <- data.frame(wavelength = seq(250, 850, by = 10))
#' reference$value <- planck(reference$wavelength, 3100)
#' reference$rsd <- 0.03
#' wl <- seq(260, 840, by = 0.1)
#' true_response <- exp(-((wl - 500) / 250)^2)
#' set.seed(1)
#' lamp <- 2e4 * true_response * planck(wl, 3100) / planck(500, 3100) + rnorm(length(wl), sd = 20)
#' names(lamp) <- wl
#' response <- response_calibration(lamp, reference)
#'
#' # a spectrum with two lines of equal emission, at 300 and 650 nm
#' emission <- 1000 * exp(-(wl - 300)^2 / 0.01) + 1000 * exp(-(wl - 650)^2 / 0.01)
#' spectrum <- stats::setNames(true_response * emission + rnorm(length(wl)), wl)
#' lines <- line_intensities(spectrum, c(a = 300, b = 650))
#' lines$intensity # unequal: the response differs
#' corrected <- correct_response(lines, response)
#' corrected[c("line", "intensity", "intensity_se", "response")]
#'
#' # or the whole spectrum
#' corrected_spectrum <- correct_response(spectrum, response)
correct_response <- function(x, response) {
if (!inherits(response, "specproc_response")) {
stop("'response' must be returned by response_calibration().", call. = FALSE)
}
r <- response$response
if (is.data.frame(x) && all(c("wavelength", "intensity") %in% names(x))) {
return(correct_line_response(x, r))
}
nms <- if (is.null(dim(x))) names(x) else colnames(x)
if (is.null(nms)) {
stop("'x' must be spectra with wavelengths as names, or a table of lines.", call. = FALSE)
}
wl <- suppressWarnings(as.numeric(nms))
channel <- !is.na(wl)
if (is.data.frame(x)) channel <- channel & vapply(x, is.numeric, logical(1))
if (sum(channel) != nrow(r) || max(abs(wl[channel] - r$wavelength)) > 1e-6) {
stop("'x' does not have the wavelength axis of the response.", call. = FALSE)
}
if (is.null(dim(x))) {
x[channel] <- x[channel] / r$response
} else if (is.matrix(x)) {
x[, channel] <- sweep(x[, channel, drop = FALSE], 2, r$response, "/")
} else {
cols <- which(channel)
for (j in seq_along(cols)) x[[cols[j]]] <- x[[cols[j]]] / r$response[j]
}
x
}
#' @title Plot a Spectral Response
#'
#' @description
#' Plots the relative spectral response measured by
#' [response_calibration()], with a band of plus or minus one standard
#' uncertainty.
#'
#' @param object The result of [response_calibration()].
#' @param title The plot title. By default, the number of lamp spectra and
#' the median uncertainty.
#' @param wavelength Optional (`NULL` by default). The wavelength range
#' `c(min, max)` to draw, in nm.
#' @param xlab,ylab The axis titles. Default are `"Wavelength (nm)"` and
#' `"Relative response"`.
#' @param base_size The size of the text, in points. Default is 11.
#' @param linewidth The width of the response line. Default is 0.6.
#'
#' @return A ggplot object.
#' @seealso [response_calibration()]
#' @export plot_response
#'
#' @examples
#' planck <- function(wl, t) 1 / (wl^5 * (exp(1.4388e7 / (wl * t)) - 1))
#' reference <- data.frame(wavelength = seq(250, 850, by = 10))
#' reference$value <- planck(reference$wavelength, 3100)
#' reference$rsd <- 0.03
#' wl <- seq(260, 840, by = 0.1)
#' set.seed(1)
#' lamp <- 2e4 * exp(-((wl - 500) / 250)^2) * planck(wl, 3100) / planck(500, 3100) +
#' rnorm(length(wl), sd = 20)
#' names(lamp) <- wl
#' plot_response(response_calibration(lamp, reference))
plot_response <- function(object, title = NULL, wavelength = NULL, xlab = "Wavelength (nm)",
ylab = "Relative response", base_size = 11, linewidth = 0.6) {
if (!inherits(object, "specproc_response")) {
stop("'object' must be returned by response_calibration().", call. = FALSE)
}
check_plot_text(title, xlab, ylab, base_size)
check_size(linewidth, "linewidth")
window <- wavelength_window(wavelength)
d <- as.data.frame(object$response)
d <- d[!is.na(d$response) & in_window(d$wavelength, window), , drop = FALSE]
if (nrow(d) < 2) {
stop("Fewer than 2 channels with a response in the 'wavelength' range.", call. = FALSE)
}
d$lower <- d$response * (1 - d$rsd)
d$upper <- d$response * (1 + d$rsd)
d$segment <- factor(d$segment)
if (is.null(title)) {
title <- sprintf("Spectral response: %d lamp spectr%s, median uncertainty %s%%",
object$n_spectra, if (object$n_spectra == 1) "um" else "a",
format(signif(100 * stats::median(d$rsd), 2)))
}
p <- ggplot2::ggplot(d, ggplot2::aes(x = .data$wavelength, group = .data$segment)) +
ggplot2::geom_ribbon(ggplot2::aes(ymin = .data$lower, ymax = .data$upper), fill = "grey75") +
ggplot2::geom_line(ggplot2::aes(y = .data$response), colour = "#1b9e77",
linewidth = linewidth) +
ggplot2::labs(x = xlab, y = ylab, title = title) +
ggplot2::theme_bw(base_size = base_size)
finish_title(p)
}
#' @export
print.specproc_response <- function(x, ...) {
r <- x$response
ok <- !is.na(r$response)
cat("Spectral response from ", x$n_spectra, " lamp spectr", if (x$n_spectra == 1) "um" else "a",
if (x$smooth > 0) paste0(" (smoothed over ", format(x$smooth), " nm)"), "\n\n", sep = "")
cat("Channels: ", nrow(r), " (", format(round(min(r$wavelength), 1)), " to ",
format(round(max(r$wavelength), 1)), " nm, ", length(unique(r$segment)), " segment",
if (length(unique(r$segment)) > 1) "s", ")\n", sep = "")
pct <- function(v) paste0(format(signif(100 * v, 2)), "%")
if (any(ok)) {
cat("Uncertainty: median ", pct(stats::median(r$rsd[ok])), " (", pct(min(r$rsd[ok])), " to ",
pct(max(r$rsd[ok])), "); reference ", pct(stats::median(r$reference_rsd[ok])),
", measurement ", pct(stats::median(r$measurement_rsd[ok])), " (medians)\n", sep = "")
}
if (nrow(x$exclude) > 0) {
cat("Excluded: ", paste(format(x$exclude[, 1]), "to", format(x$exclude[, 2]), "nm",
collapse = ", "), "\n", sep = "")
}
if (any(!ok)) {
cat("No response: ", sum(!ok), " channel(s) (the reference covers ",
paste(format(x$range), collapse = " to "), " nm)\n", sep = "")
}
invisible(x)
}
# ---- internals ---------------------------------------------------------------
check_response_reference <- function(reference) {
if (!is.data.frame(reference) || !all(c("wavelength", "value") %in% names(reference))) {
stop("'reference' must be a data frame with columns `wavelength` and `value`.", call. = FALSE)
}
ok <- is.numeric(reference$wavelength) && is.numeric(reference$value) &&
!anyNA(reference$wavelength) && !anyNA(reference$value) && all(reference$value > 0)
if (!ok || nrow(reference) < 2 || anyDuplicated(reference$wavelength)) {
stop("'reference' needs at least 2 distinct wavelengths with positive values.", call. = FALSE)
}
if ("rsd" %in% names(reference) &&
(!is.numeric(reference$rsd) || anyNA(reference$rsd) || any(reference$rsd < 0))) {
stop("The column `rsd` of 'reference' must contain non-negative numbers.", call. = FALSE)
}
reference[order(reference$wavelength), , drop = FALSE]
}
# Certified values at the wavelengths wl: monotone cubic spline of their
# logarithm, NA outside their range.
reference_values <- function(reference, wl) {
f <- stats::splinefun(reference$wavelength, log(reference$value), method = "monoH.FC")
out <- exp(f(wl))
out[wl < min(reference$wavelength) | wl > max(reference$wavelength)] <- NA
out
}
# Wavelength ranges as a two-column matrix.
exclude_ranges <- function(exclude) {
if (is.null(exclude)) return(matrix(numeric(), 0, 2))
m <- if (is.matrix(exclude)) exclude else if (is.list(exclude)) do.call(rbind, exclude) else
matrix(exclude, ncol = 2, byrow = TRUE)
if (!is.numeric(m) || ncol(m) != 2 || anyNA(m) || any(m[, 1] >= m[, 2])) {
stop("'exclude' must give wavelength ranges (from, to) with from < to.", call. = FALSE)
}
unname(m)
}
# Local linear fit of v over `width` nm around each channel, within each
# segment, with the number of channels of each window; the channels without
# a value are interpolated from their neighbors.
smooth_response <- function(v, wl, seg, width) {
fit <- n <- rep(NA_real_, length(v))
for (s in unique(seg)) {
idx <- which(seg == s)
idx <- idx[order(wl[idx])] # overlapping detectors in one segment
one <- local_linear(v[idx], wl[idx], width)
fit[idx] <- one$fit
n[idx] <- one$n
}
list(fit = fill_segments(fit, wl, seg), n = n)
}
local_linear <- function(v, w, width) {
ok <- !is.na(v)
if (width <= 0) return(list(fit = v, n = as.numeric(ok)))
w0 <- w - mean(w)
total <- function(a) c(0, cumsum(ifelse(ok, a, 0)))
window <- function(s) s[hi + 1] - s[lo + 1]
lo <- findInterval(w - width / 2, w, left.open = TRUE) # channels before the window
hi <- findInterval(w + width / 2, w) # channels up to its end
n <- window(total(1))
sw <- window(total(w0))
sv <- window(total(v))
sww <- window(total(w0^2))
swv <- window(total(w0 * v))
var <- sww - sw^2 / n
slope <- ifelse(n >= 2 & var > 1e-12, (swv - sw * sv / n) / var, 0)
fit <- ifelse(n > 0, sv / n + slope * (w0 - sw / n), NA_real_)
list(fit = fit, n = n)
}
# Linear interpolation of the missing values of v within each segment.
fill_segments <- function(v, wl, seg) {
for (s in unique(seg)) {
idx <- which(seg == s)
ok <- !is.na(v[idx])
if (any(ok) && !all(ok)) {
v[idx] <- if (sum(ok) == 1) v[idx][ok] else
stats::approx(wl[idx][ok], v[idx][ok], xout = wl[idx], rule = 2, ties = mean)$y
}
}
v
}
# Divides the intensities of a table of lines by the response at their peak.
correct_line_response <- function(lines, r) {
at <- lines$wavelength
if ("peak_wavelength" %in% names(lines)) {
at <- ifelse(is.na(lines$peak_wavelength), at, lines$peak_wavelength)
}
idx <- vapply(at, function(w) which.min(abs(r$wavelength - w)), integer(1))
resp <- r$response[idx]
rsd <- r$rsd[idx]
far <- abs(r$wavelength[idx] - at) > 1
resp[far] <- NA
rsd[far] <- NA
if (anyNA(resp)) {
warning(sum(is.na(resp)), " line(s) have no response at their wavelength; their ",
"intensity is NA.", call. = FALSE)
}
se_rel <- if ("intensity_se" %in% names(lines)) lines$intensity_se / lines$intensity else 0
lines$intensity <- lines$intensity / resp
if ("height" %in% names(lines)) lines$height <- lines$height / resp
lines$intensity_se <- lines$intensity * sqrt(se_rel^2 + rsd^2)
lines$response <- resp
lines
}
Any scripts or data that you put into this service are public.