Preprocessing

This vignette provides an overview of the preprocessing tools available for data in MetaboSet objects.

Drift correction

Untargeted LC-MS metabolomics data usually suffers from signal intensity drift, i.e. systematic variation in the signal intensity as a function of injection order. Features of a dataset often have very different drift patterns. This means that to correct for the drift, we need a tool that can adapt to different drift patterns.

Most drift correction methods follow a simple logic:
1. Model the systematic drift using QC samples.
2. Correct the abundances of all the samples by reversing the modeled effect of the drift.

The approach used in this package models the drift separately for each feature, using cubic spline regression. The R function \code{smooth.spline} performs the modelling. The function only needs one hyperparameter, a so-called smoothing parameter. The smoothing parameter controls the "wiggliness" of the function: the larger the smoothing parameter, the closer the spline is to a linear function. By default, the smoothing parameter is chosen automatically using cross validation, and is allowed to vary between 0.5 and 1.5, where 1.5 is large enough to guarantee that the curve is practically linear. This interval seems to work well in practice.

We run the drift correction on log-transformed data. Log-transformed data follows the assumptions of the drift function model better, and also avoids some practical issues. The corrected feature abundance for a sample with injection order $i$ is then computed using the following formula: $$x_{corrected}(i) = x_{original}(i) + \overline{x}{QC} - f{drift}(i)$$

The below figure illustrates the procedure.

knitr::include_graphics("drift_corr.png")

The plots on the left show the feature abundances before and after the drift correction procedure. The plots on the right show the drift correction procedure in log-transformed space. The horizontal lines show deviations of +- 2 SD from the mean for QCs and biological samples. Note how the cubic spline can fit nonlinear drift patterns quite nicely.

Code

To run drift correction on your data, use the function correct_drift. It will allow you to run drift correction, adjust parameters of the cubic spline regression and optionally plot figures such as the above for your features.

Flagging low quality compounds

LC-MS metabolomics datasets contain a significant amount of low-quality features. We use certain quality metrics to flag those features. Flagged features are not removed from the data, but the flag information is stored on the "Flag" column of the results part of the MetaboSet object. But if the features are not removed, then what difference does it make if they are flagged or not? Well, a rule of thumb about the role of flagged features is as followed: whenever you do operations that only deal with one feature at a time, flagged features are processed just like any other feature. Drift correction and log transformation are good examples of such procedure. But when you are running operations that involve multiple features at a time, such as random forest imputation, PCA visualization or any multivariate statistical models, the flagged features are silently ignored. For feature-wise statistics,the statistical model is fit for flagged features, but FDR correction is only done for non-flagged features. Most functions in the package have an argument called all_features, which is FALSE by default, and can be set to TRUE to force the function to treat all features as equal (ignore flags).

The primary function of the QC samples is to allow us to (surprise) assess the quality of the molecular features. In essence, the variation in the feature abundance of QC samples should be small. But how do we define "small"? Luckily, this question is answered in the paper by Broadhurst et.al., which provides us with five useful quality metrics.

Detection rate

First, we look at QC detection rate. This can be defined as simply the number of detected QC samples divided by the number of expected QC samples for a given metabolite, expressed as a percentage. If a signal can't be reliably detected in QC samples, it is regarded as noise. The paper suggests an acceptance limit of 70%, but this seems quite harsh for studies where the groups differ significantly. In these cases, some group might have metabolites that are not present in others, which means that their concentration in the QC samples might fall below the resolution of the mass spectrometer.

Additionally, it is sometimes useful to flag features that

Spread of the QCs

The actual quality metrics measure two things: internal spread of the QCs, and spread of the QCs compared to the spread of the biological samples.

Internal spread is measured with relative standard deviation (RSD), also known as coefficient of variation (CV).

$$ RSD = \frac{s_{QC}}{\bar{x}_{QC}} $$

Where $s_{QC}$ is the standard deviation of the QC samples and $\bar{x}_{QC}$ is the sample mean of the signal in the QC samples. RSD can also be replaced by a non-parametric, robust version based on the median and median absolute deviation (MAD):

$$RSD^* = \frac{1.4826 \cdot MAD_{QC}}{\text{median}(x_{QC}}$$

We are using the robust version as it is less affected by a single outlying QC and since the original RSD works best if the data is normally distributed, which is often not the case with our data.

The spread of the QC samples compared to the biological samples is measured using a metric called D-ratio:

$$D-ratio = \frac{s_{QC}}{s_{biological}}$$

Or, as before, a non-parametric, robust alternative:

$$D-ratio* = \frac{MAD_{QC}}{MAD_{biological}}$$

We are using the robust alternative of D-ratio as well, for the same reasons as above.

Limits used in filtering

The recommended limits for RSD and D-ratio are 0.2 and 0.4, respectively. So the signals need to score lower than the limits in both metrics to survive filtering. Since we are using the robust alternatives, we have added an additional condition: Signals with classic RSD, RSD* and basic D-ratio all below 0.1 are kept. This additional condition prevents the removal of signals with very low values in all but a few samples. These signals tend to have a very high value of D-ratio*, since the median absolute deviation of the biological samples is not affected by the large concentration in a handful of samples, causing the D-ratio* to overestimate the significance of random errors in measurements of QC samples. Thus, other quality metrics are applied, with a conservative limit of 0.1 to ensure that only good quality signals are kept this way.

Use in overall quality assessment

The distribution of the quality metrics across signals can be used as a good quality metric of a study. In the future, some key statistics, for example the mean, median and maybe some other quantiles could be reported for each statistic. This would yield us a nice way of quantifying the quality of the data from any given experiment.

Code

flag_detection is used to flag features based on detection