knitr::opts_chunk$set( collapse = TRUE, comment = "#>", fig.width = 7, fig.height = 4, out.width = "100%", dpi = 96 )
Quantitative LIBS assumes that the plasma is in local thermodynamic
equilibrium (LTE), optically thin, and recorded within the linear range of
the detector. This vignette checks these conditions on the forage spectra,
measures the electron density (N_e) and the temperature (T) of the
plasma, and ends with a calibration-free estimate of the composition of the
samples. The intensities are not corrected for the spectral response of the
instrument, which needs spectra of a calibration lamp (see
response_calibration()), and the acquisition delay is not known, so the
values are indicative.
library(specProc) library(dplyr) library(tidyr) library(tibble) library(ggplot2) data("forageLIBS") spectra_id <- forageLIBS |> select(1:2) |> names() minerals <- forageLIBS |> select(3:14) |> names() spectra <- forageLIBS |> select(-all_of(c(spectra_id, minerals))) wl <- as.numeric(names(spectra)) corrected <- baseline_arpls(spectra, lambda = 1e5, max_iter = 20)$correction mean_spectrum <- colMeans(corrected)
A saturated line is clipped at the maximum count of the detector: its area
is wrong and no longer grows with concentration. saturation_summary()
lists the saturated channels. Each spectrum is the mean of 8 shots, so we
count the channels within 1% of the 16-bit limit (65535), that is,
saturated in nearly every shot:
saturated <- saturation_summary(spectra, limit = 65535, tolerance = 655) # adjacent saturated channels belong to the same line saturated$channels |> mutate(line = cumsum(c(1, diff(index) > 2))) |> group_by(line) |> summarise(wavelength = round(mean(wavelength), 2), spectra = max(n_spectra)) |> select(-line) |> arrange(desc(spectra))
The resonance lines of Ca II (393.37 and 396.85 nm), Mg II (279.55 and 280.27 nm), Ca I (422.67 nm), Na I (589.00 and 589.59 nm) and K I (769.90 nm), as well as the head of the CN molecular band at 388.3 nm, are saturated in part of the spectra: they are left out of every diagnostic below.
The hydrogen line H(\alpha) (656.28 nm) is the standard density
diagnostic: its Stark width is large and depends little on temperature
(Gigosos et al., 2003). Its profile is Lorentzian in these spectra (see
vignette("peak-fitting")), and electron_density() converts its full
width at half maximum into (N_e):
halpha <- as_tibble(as.list(mean_spectrum[wl > 653.5 & wl < 659.5])) |> peak_fit(profile = "lorentzian") width <- halpha$tidied[[1]] |> filter(term == "wL") width ne <- electron_density(width$estimate, method = "halpha") ne
The fit also gives the standard error of the width.
electron_density_uncertainty() propagates it to the density, together
with the uncertainty of the width law of H(\alpha) itself, taken here as
10% for illustration:
ne_u <- electron_density_uncertainty(width$estimate, width$std.error, method = "halpha", reference_rsd = 0.1) ne_u
The width of H(\alpha) grows as (N_e^{0.68}), so a relative error of the
width is amplified about 1.5 times in the density. The relative standard
uncertainty of the density, r round(100 * ne_u$rsd)%, comes mostly from the
width law: the fit error alone is r round(100 * width$std.error / width$estimate, 1)%
of the width.
For other lines, stark_width() gives the width expected at a density from
the Stark parameters of the STARK-B database (starkb_lines()), and
self_absorption() compares it with the measured width.
A Boltzmann plot needs, for each line, the transition probability
(A_{ki}), the statistical weight (g_k) and the energy (E_k) of the upper
level. nist_lines() retrieves them from the NIST Atomic Spectra Database,
and nist_ionization_energy() the ionization energies:
ca_lines <- nist_lines(c("Ca I", "Ca II"), wavelength = c(310, 660)) nist_ionization_energy("Ca I")
So that this vignette runs offline, the calcium lines used below are included (NIST ASD, Kramida et al.). The accuracy grades of the transition probabilities (C: ≤ 25%, D: ≤ 50%) limit the precision of the temperature:
atomic <- tribble( ~species, ~stage, ~wavelength, ~Aki, ~accuracy, ~Ek, ~gk, "Ca I", 1, 428.301, 4.34e7, "C+", 4.779784, 5, "Ca I", 1, 430.253, 1.36e8, "C+", 4.779784, 5, "Ca I", 1, 431.865, 7.40e7, "C+", 4.7690284, 3, "Ca I", 1, 443.496, 6.70e7, "C", 4.680635, 5, "Ca I", 1, 445.478, 8.70e7, "C", 4.681327, 7, "Ca I", 1, 558.876, 4.90e7, "D", 4.7435268, 7, "Ca I", 1, 610.272, 9.60e6, "C", 3.910399, 3, "Ca I", 1, 612.222, 2.87e7, "C", 3.910399, 3, "Ca I", 1, 616.217, 4.77e7, "C", 3.910399, 3, "Ca I", 1, 643.907, 5.30e7, "D", 4.450647, 9, "Ca I", 1, 646.257, 4.70e7, "D", 4.4409544, 7, "Ca I", 1, 649.378, 4.40e7, "D", 4.4300117, 5, "Ca II", 2, 315.887, 3.10e8, "C", 7.047169, 4, "Ca II", 2, 317.933, 3.60e8, "C", 7.049551, 6, "Ca II", 2, 370.603, 8.80e7, "C", 6.467875, 2, "Ca II", 2, 373.690, 1.70e8, "C", 6.467875, 2 ) ionization_energy <- 6.1131549 # Ca I, eV
line_intensities() finds each line within 0.2 nm of its tabulated
wavelength (search), which absorbs the offset of the wavelength axis, and
integrates it over ±0.15 nm (half_width). It keeps the columns of the
table of lines, so the result goes directly into a Boltzmann plot:
lines <- line_intensities(mean_spectrum, atomic) lines |> select(species, wavelength, peak_wavelength, intensity, snr)
neutral <- boltzmann(filter(lines, stage == 1)) neutral
The Ca I lines come from upper levels between 3.9 and 4.8 eV. Over such a narrow range, the slope is sensitive to the errors of a few intensities or transition probabilities. The Saha equation places the lines of the ion on the same plot, using the electron density, and extends the range of energies to about 13 eV:
saha <- saha_boltzmann(lines, ionization_energy = ionization_energy, electron_density = ne) saha
plot_boltzmann(saha)
The Saha-Boltzmann temperature, r round(saha$temperature, -2) K, is far
better determined than the Boltzmann temperature of the neutral lines
alone (r round(neutral$temperature, -2) ±
r round(neutral$temperature_se, -1) K). Its standard error,
r round(saha$temperature_se, -1) K, reflects the scatter of the points.
tidy() lists the estimates of a fit, here with their 95% confidence
intervals (that of the temperature is the interval of the slope,
transformed), and glance() summarizes the fit in one row, to compare
plots or bind the results of several spectra:
tidy(saha, conf.int = TRUE) glance(saha)
The lines of a Boltzmann plot can also be chosen interactively:
line_finder() measures the lines picked on a spectrum, fits their
Boltzmann plot as they are added, and leaves out the points clicked on the
plot.
The points scatter because of the errors of the intensities and of the
transition probabilities. The accuracy grades of the NIST database (column
accuracy) bound the standard uncertainties of the transition
probabilities: 18% for grade C+, 25% for C and 50% for D (see
aki_uncertainty()). line_intensities() gives the standard error of each
intensity due to the noise of the spectrum (intensity_se).
plasma_uncertainty() propagates these uncertainties, and that of the
electron density, to the temperature: it repeats the fit with inputs
perturbed within their uncertainties (Monte Carlo simulation):
set.seed(1) saha_u <- plasma_uncertainty(saha, electron_density_se = ne_u$se) saha_u
The intensities of the mean spectrum are measured to about 1%, so the
transition probabilities and the electron density, which contribute about
equally, make up the interval,
r round(saha_u$estimates$lower, -1) to r round(saha_u$estimates$upper, -1)
K. The reduced chi-squared compares the scatter of the points with their
stated uncertainties. At r round(saha_u$chi_squared, 1), the scatter of
the points is about r format(round(sqrt(saha_u$chi_squared), 1), nsmall = 1)
times what these uncertainties explain: other errors are at work, such as
the spectral response of the
instrument, which is not corrected here, or self-absorption. The interval
does not include them, so it is a lower bound of the uncertainty.
The McWhirter criterion gives the minimum electron density for collisions to dominate the population of the levels. The largest energy gap here is the first excited level of Ca II, 3.15 eV above the ground state:
lte <- mcwhirter_criterion(temperature = saha$temperature, delta_e = 3.15, electron_density = ne) lte
The measured density is r round(ne / lte$minimum_density) times the
minimum. The criterion is necessary but not sufficient: in a transient,
inhomogeneous plasma, LTE also requires the equilibration to be faster
than the changes of the plasma (Cristoforetti et al., 2010).
lte_criteria() computes the time the plasma takes to relax to
equilibrium, through the resonance transition of Ca II (393.37 nm,
oscillator strength 0.68), and the distance an atom diffuses meanwhile,
here compared with 1 mm, the order of the size of the plasma:
criteria <- lte_criteria(temperature = saha$temperature, electron_density = ne, delta_e = 3.15, oscillator_strength = 0.68, mass = "Ca", length_scale = 1) criteria |> select(mcwhirter, relaxation_time, diffusion_length, length_ratio, spatial)
The relaxation time, r signif(1e9 * criteria$relaxation_time, 2) ns, is
far shorter than the microseconds over which a LIBS plasma cools, and the
diffusion length, r signif(1e3 * criteria$diffusion_length, 2) µm, far
shorter than the plasma: the spatial criterion is satisfied. The temporal
criterion needs the temperature and density at several delays (time),
which these spectra, recorded at one delay, do not have.
Resonance lines end on the ground state, which is heavily populated, so
the plasma reabsorbs part of their light. correct_self_absorption()
compares each line with a reference line of the same species that is
little absorbed, by default the one of smallest optical depth (Sun and Yu,
2009). The potassium lines:
k_lines <- tribble( ~species, ~wavelength, ~Aki, ~Ei, ~Ek, ~gk, "K I", 404.414, 1.150e+06, 0, 3.064907, 4, "K I", 404.721, 1.070e+06, 0, 3.062581, 2, "K I", 691.108, 2.500e+06, 1.609958, 3.403454, 2, "K I", 693.877, 4.956e+06, 1.617113, 3.403454, 2, "K I", 766.490, 3.779e+07, 0, 1.617113, 4, "K I", 769.896, 3.734e+07, 0, 1.609958, 2 ) line_intensities(mean_spectrum, k_lines) |> correct_self_absorption(temperature = saha$temperature) |> select(wavelength, measured_intensity, SA, reference)
The self-absorption coefficient SA is the fraction of the intensity that
escapes the plasma. The resonance doublet at 766.49 and 769.90 nm keeps only
a small fraction: the 766.49 nm line has twice the (g_k A_{ki}) of the
769.90 nm line, so it should be twice as intense, yet it is weaker. These
lines curve the calibration of vignette("calibration") and are unsuitable
for the calibration-free analysis below. The coefficients of the weaker
lines, between about 0.6 and 1.07 (hence the warning), show the precision
of the method: it inherits the noise of the reference line and the errors
of the transition probabilities.
curve_of_growth() needs no reference line. It fits the curve of growth of
a homogeneous plasma in LTE to the lines of one or several species at once,
the model of the C-sigma graphs of Aragón and Aguilera (2014): the
intensity of a line grows as its optical depth while the plasma is thin,
then more slowly as its center saturates. With one column density per
species, the fit gives the optical depth of every line, its
self-absorption coefficient and its optically thin intensity, ready for
boltzmann() or cf_libs(). It needs the Stark widths of the lines, which
set how fast they saturate: stark_width() gives them from the STARK-B
database (starkb_lines()) at the electron density measured above, and
plot_curve_of_growth() draws the fit.
Calibration-free LIBS (CF-LIBS; Ciucci et al., 1999) derives the
composition of a sample from its line intensities alone. In LTE, the
Boltzmann plots of all species share the same slope, (-1/k_B T), and the
intercept of each plot gives the density of its species. cf_libs() fits
the plots together, adds the ionization stages that are not observed with
the Saha equation, and normalizes the densities of the elements to
fractions that sum to one.
The lines must be unsaturated and optically thin, which rules out the
resonance lines. cf_libs() also needs the partition functions of the
species, which it computes from the energy levels of nist_levels(), and
the ionization energies:
species <- c("Ca I", "Ca II", "Mg I", "Mg II", "K I", "K II", "Na I", "Na II") levels <- nist_levels(species) ionization <- nist_ionization_energy(c("Ca I", "Mg I", "K I", "Na I"))
To run offline, the atomic data and the partition functions (from
partition_function(), tabulated from 6000 to 12000 K) are included here:
cf_atomic <- tribble( ~species, ~wavelength, ~Aki, ~accuracy, ~Ek, ~gk, "Ca I", 428.301, 4.340e+07, "C+", 4.779784, 5, "Ca I", 430.253, 1.360e+08, "C+", 4.779784, 5, "Ca I", 431.865, 7.400e+07, "C+", 4.769028, 3, "Ca I", 443.496, 6.700e+07, "C", 4.680635, 5, "Ca I", 445.478, 8.700e+07, "C", 4.681327, 7, "Ca I", 616.217, 4.770e+07, "C", 3.910399, 3, "Ca II", 315.887, 3.100e+08, "C", 7.047169, 4, "Ca II", 317.933, 3.600e+08, "C", 7.049551, 6, "Ca II", 370.603, 8.800e+07, "C", 6.467875, 2, "Ca II", 373.690, 1.700e+08, "C", 6.467875, 2, "Mg I", 383.829, 1.610e+08, "B+", 5.945916, 7, "Mg I", 516.732, 1.130e+07, "B+", 5.107827, 3, "Mg I", 517.268, 3.370e+07, "B+", 5.107827, 3, "Mg I", 518.360, 5.610e+07, "A", 5.107827, 3, "Mg II", 279.078, 4.010e+08, "A", 8.863762, 4, "Mg II", 279.800, 4.790e+08, "A", 8.863654, 6, "Mg II", 292.863, 1.150e+08, "A", 8.654711, 2, "Mg II", 293.651, 2.300e+08, "A", 8.654711, 2, "K I", 404.414, 1.150e+06, "A", 3.064907, 4, "K I", 404.721, 1.070e+06, "B+", 3.062581, 2, "K I", 691.108, 2.500e+06, "AA", 3.403454, 2, "K I", 693.877, 4.956e+06, "AA", 3.403454, 2, "Na I", 818.326, 4.290e+07, "A+", 3.616977, 4, "Na I", 819.482, 5.140e+07, "A+", 3.616971, 6 ) cf_ionization <- c(Ca = 6.113155, Mg = 7.646236, K = 4.340664, Na = 5.139077) partition_table <- tribble( ~species, ~`6000`, ~`7000`, ~`8000`, ~`9000`, ~`10000`, ~`11000`, ~`12000`, "Ca I", 1.407, 1.798, 2.401, 3.281, 4.499, 6.184, 8.356, "Ca II", 2.389, 2.633, 2.917, 3.229, 3.560, 3.905, 4.262, "Mg I", 1.049, 1.109, 1.207, 1.362, 1.597, 1.959, 2.463, "Mg II", 2.001, 2.004, 2.010, 2.020, 2.036, 2.058, 2.087, "K I", 2.488, 3.000, 3.858, 5.078, 6.743, 8.938, 11.503, "K II", 1, 1, 1, 1, 1, 1, 1, "Na I", 2.166, 2.401, 2.868, 3.635, 4.779, 6.548, 8.724, "Na II", 1, 1, 1, 1, 1, 1, 1 ) # log-linear interpolation of the table, in the form cf_libs() accepts partition <- function(species, temperature) { grid <- as.numeric(names(partition_table)[-1]) vapply(species, function(s) { u <- unlist(partition_table[partition_table$species == s, -1]) exp(stats::approx(grid, log(u), xout = temperature)$y) }, numeric(1)) }
With partition = NULL (the default), cf_libs() downloads the levels
and computes the partition functions itself. On the mean spectrum:
cf_mean <- line_intensities(mean_spectrum, cf_atomic) |> cf_libs(electron_density = ne, partition = partition, ionization_energy = cf_ionization) cf_mean
plot_boltzmann(cf_mean)
The four Saha-Boltzmann plots are parallel within the scatter of their points. The forage is mostly C, H, O and N, whose lines are not used here, so the fractions describe the relative composition of Ca, Mg, K and Na, which we compare with the laboratory values on the same basis:
laboratory <- forageLIBS |> summarise(across(c(Ca, Mg, K, Na), \(v) mean(v, na.rm = TRUE))) |> unlist() cf_mean$composition |> select(element, cf_libs = mass_fraction) |> mutate(laboratory = laboratory[element] / sum(laboratory), ratio = cf_libs / laboratory)
ratios <- cf_mean$composition |> mutate(laboratory = laboratory[element] / sum(laboratory), ratio = mass_fraction / laboratory) ratio_of <- \(e) round(ratios$ratio[ratios$element == e], 1)
Without any calibration, the composition is of the right order: the ratios
to the laboratory values are r ratio_of("Ca") for Ca,
r ratio_of("Mg") for Mg, r ratio_of("K") for K and r ratio_of("Na")
for Na. The main sources of bias are the spectral response of the
instrument, which weights the lines of each element differently, the
uncertain transition probabilities of the Ca lines, and the weak Na lines.
With one standard of known composition, the remaining bias can be measured
and removed for each element.
plasma_uncertainty() also repeats the calibration-free analysis with
perturbed inputs:
set.seed(1) cf_u <- plasma_uncertainty(cf_mean, n = 500, electron_density_se = ne_u$se) cf_u
The intervals are narrow, a few percent of each fraction. The reduced
chi-squared, r round(cf_u$chi_squared), shows why they are misleading:
the K, Mg and Na lines have accurate transition probabilities (grades AA to
B+) and intensities measured to a few percent, so their scatter about the parallel
plots comes from other errors, such as the spectral response, self-absorption
or departures from LTE. These systematic errors, not the random ones, limit
the accuracy of the composition, as the comparison with the laboratory
values shows.
plasma_uncertainty(). A
reduced chi-squared well above 1 means that other errors dominate, and
the interval is too narrow.Any scripts or data that you put into this service are public.