Description Arguments Value Methods See Also
Calibrate peaks of a xcmsSet via a set of known masses
| object | a  | 
| calibrants | a vector or a list of vectors with reference m/z-values | 
| method | the used calibrating-method, see below | 
| mzppm | the relative error used for matching peaks in ppm (parts per million) | 
| mzabs | the absolute error used for matching peaks in Da | 
| neighbours | the number of neighbours from wich the one with the highest intensity is used (instead of the nearest) | 
| plotres | can be set to TRUE if wanted a result-plot showing the found m/z with the distances and the regression | 
| object | a  | 
| calibrants | for each sample different calibrants can be used, if a list of m/z-vectors is given. The length of the list must be the same as the number of samples, alternatively a single vector of masses can be given which is used for all samples. | 
| method | "shift" for shifting each m/z, "linear" does a linear regression and adds a linear term to each m/z. "edgeshift" does a linear regression within the range of the mz-calibrants and a shift outside. | 
    calibrate(object, calibrants,method="linear",
    mzabs=0.0001, mzppm=5,
    neighbours=3, plotres=FALSE)
  
Add the following code to your website.
For more information on customizing the embed code, read Embedding Snippets.